Specifications: | |
Infusion Mode | Rate mode, Volume-Time, Rate-Time, Dosage |
IV set | 4 preset brands and extensive |
Display | 3.5 inches TFT color touch screen, resolution:480*272
9 adjustable brightness levels |
Flow Rate Range
Increment |
0.10 ~ 2000 ml/h |
0.01ml/h (0.10-99.99ml/h)
0.1 mlh/h(100.0-999.9ml/h) 1 ml/h(1000-2000ml/h) |
|
VTBI Range | 0.10~ 9999 ml |
Flow Rate Accuracy | ≤ ± 2% |
Bolus Mode | Automatical and Manual |
Bolus Rate | 0.10 ~ 2000 ml/h |
Auto Bolus VTBI | 0.10-5 ml, at 0.1ml/h increment |
KVO Rate | 0.10-5 ml/h, at 0.1ml/h increment |
Occlusion Pressure | Dynamic Pressure System with upstream/downstream occlusion detectors |
11 sensitivities setting, 10 ~ 135KPa. | |
Drug Library | Programmable Drug Library |
Event Log | over 2000 logs |
Night Mode | Programmable and auto shift. |
Parameter Adjustable
Online |
Change parameter without stop |
Anti-Bolus Function | YES, release 70% pressure of the peak value |
Key Lock | YES |
Alarm | 3 alarm grades, 5 adjustable sound volume settings |
Data Port | P1 |
Dual CPUs | YES |
Power Source | AC: 100V ~ 240V, 50Hz/60Hz, 35VA |
Battery | Rechargeable Lithium-ion battery: working time: ≥ 8 hours at 25ml/h |
Classification | Class I, CF, IPX3 |
Dimensions and N.W. | 280mm x 170mm x 80mm; 1.8kg |
Packing: | |
Packing: | 414x238x154mm; 2.8kg |
Warranty: | 18 months |
i7 PLUS Infusion Pump
Event Log:
Event log volume: 2000 cases
Event content: Automatically record alarm information and operation event, including start injecting, injection completed, rate modification, injection volume etc.
Data Communication:
Data transmission interface: RS232; Optional WiFi
Infusion workstation: It can work together with other pumps as part of an infusion workstation.
Type of alarm: Three grade (High, Medium and Low) audible and visible alarm with text prompts; 9 levels adjustable alarm beep sound.
Alarm content:
• AC power interruption
• DC power interruption
• Low battery
• Depleted battery
• Forget reminder
• Infusion near end
• Infusion complete
• Flow rate
• Door open
• System exception
• Upstream Occlusion
• Bubble detection
Bubble detection alarm: A single bubble alarm for 25ul, 100ul, 250ul. Cumulated bubble alarm adjustable for 300ul/15min, 500ul/15min
Event Log:
Event log volume: 2000 cases
Event content: Automatically record alarm information and operation event, including start injecting, injection completed, rate modification, injection volume etc.
Data Communication:
Data transmission interface: RS232; Optional WiFi
Infusion workstation: It can work together with other pumps as part of an infusion workstation.
Type of alarm: Three grade (High, Medium and Low) audible and visible alarm with text prompts; 9 levels adjustable alarm beep sound.
Alarm content:
• AC power interruption
• DC power interruption
• Low battery
• Depleted battery
• Forget reminder
• Infusion near end
• Infusion complete
• Flow rate
• Door open
• System exception
• Upstream Occlusion
• Bubble detection
Bubble detection alarm: A single bubble alarm for 25ul, 100ul, 250ul. Cumulated bubble alarm adjustable for 300ul/15min, 500ul/15min